Your Position: Home > Support > Service Center

Motor wiring and rotating direction of ZYD two-stage vacuum oil filter SV Laibao

2019/8/1??????view:

  The maximum exhaust pressure shall not be greater than 1.15 bar (absolute pressure). Strictly prevent any blockage in the exhaust pipe. Exhaust filters, accessories and pipes must be estimated according to maximum flow rate. The maximum flow rate is equivalent to the pumping speed of the pump.

  Electrical Connection of ZYD Dual Stage Oil Filter SV Laibao

  1. Wiring warning for motor of vacuum oil filter:

  The input power of the pump must be cut off before the connection of the motor or the replacement of the connection. Electrical wiring must be carried out by qualified electricians in accordance with the application safety regulations.

  The motor connection is connected to the correct supply voltage through the connector in the junction box. See the wiring diagram on the motor or in the wiring box. Correct electrical connection must use appropriate motor protection switch. Set the switch according to the rated value on the motor nameplate.

  2. Cautions for connection of motor of vacuum oil filter:

  After the connection of the motor and whenever the connection is changed, the rotation direction of the motor should be checked. The direction of rotation of the motor should be the same as the direction arrow on the motor cover. When checking, the intake should be opened. If the motor rotates in the wrong direction, oil may be ejected from the intake port. Pressure in the vacuum system may increase.

  Do not use motor fans to check the direction of rotation. Because the motor fan rotates too fast to judge the direction of rotation; when the parking slows down, the fan will keep the original direction of rotation.

  Start the pump briefly, check whether the intake is sucked into the feeling, and then immediately stop the pump. When the pump steers correctly, you should feel sucked into the intake port.

  If you need to change the direction of rotation, cut off the power supply, and exchange two-phase wiring.

  We recommend that the rotation direction be checked with a phase sequence indicator.

  Long-term operation of the motor in the opposite direction will damage the pump!

  Optional accessory overheat protection switch is provided. If the power supply of the pump is cut off beyond the specified temperature, the alarm signal can be triggered at the same time. The overheat protection switch can be installed on the screw plug position (6/26) of the fuel tank (6/27). No screw plug is installed. The 6.35mm FASTON clamp is used for electrical wiring.

  Power supply: 25VAC (maximum) 60VDC (maximum)

  3. Motor switch warning of vacuum oil filter

  If the safety switch or electrical failure will turn off the pump, the re-start of the pump can only be done manually.


主站蜘蛛池模板: 日韩欧美在线视频观看 | 看片午夜| 韩国美女福利专区一区二区 | 久久精品国产99久久3d动漫 | 可播放的欧美男男videos | 亚洲午夜免费视频 | 久久精品视频久久 | 91麻豆精品福利在线观看 | 色老头久久久久久久久久 | 亚洲影院在线播放 | 一区二区三区四区在线不卡高清 | 亚洲欧美日韩国产精品第不页 | 男人和女人做免费做爽爽视频 | 日韩第一页 | 成人免费观看视频 | 日韩欧美在线不卡 | 美国十次怡红院 | aaa亚洲| 色综合精品久久久久久久 | 亚洲另类伦春色综合妖色成人网 | 国产亚洲精品久久综合影院 | 国产免费啪视频观看网站 | 香蕉视频一区二区 | 日日夜夜天天干干 | 亚洲精品一区二区中文 | 国产三级在线观看视频不卡 | 日本性插 | 亚洲视频在线一区二区三区 | 天天色综合6 | 国产精品青青青高清在线 | 久久精品国产只有精品66 | 在线观看老湿机免费福利视频 | 狠狠操夜夜爽 | 国产精品99久久免费观看 | 日韩性网 | 色狠狠一区二区三区香蕉蜜桃 | 斗罗大陆动漫在线观看免费樱花动漫 | 天天插天天操天天干 | 精品福利一区二区在线观看 | 日韩免费一区二区三区在线 | 天天综合天天做 |